Hampshire Controls CA5000-C CO2 backup alarm system

  • Model:CA5000-C
  • Temperature range: -200 °C to +50 °C
  • Liquid: CO2

Hampshire Controls CA5000-C CO₂ backup alarm system

Hampshire Controls CA5000-C CO₂ backup alarm system is specially designed to keep critical samples safe during power outages or mechanical failures of ultra-low temperature freezers. With its automatic CO₂ injection system and advanced alarm functionality, this system provides maximum protection and peace of mind in laboratories, hospitals, biobanks and pharmaceutical environments.

Why choose Hampshire Controls’ CA5000-C?

  • Reliable CO₂ backup: automatically injects liquid CO₂ as soon as the temperature rises above the set limit.
  • Autonomous operation in case of power failure: thanks to the built-in battery with a backup duration of up to 48 hours, the system remains active and alerts personnel in case of failure.
  • High precision and flexibility: precise temperature sensor (Type T) with a measurement range of -40°C to -150°C and programmable alarm limits.
  • Clear notifications: visual and audible alarms, with adjustable alarm delays, and relay outputs for connection to central reporting systems.
  • Space-saving design: compact modules that are easy to install, whether for wall mounting or directly on or next to freezers.

Optimal protection of your samples

The CA5000-C CO₂ backup alarm system from Hampshire Controls is essential in situations where sample integrity must never be compromised. In the event of a sudden failure, the system immediately takes over protection with CO₂ injection and alarm. This keeps the internal temperature within safe limits, even without mains power.

Plug and play installation

The system is easy to install and requires minimal technical knowledge. Plug in the sensor, set the desired temperature and alarm delays, and you’re instantly protected. A 22.7 kg CO₂ cylinder guarantees up to 10 hours of backup at a freezer setting of -60°C.

Interested? Contact us or download the brochure to learn more about the CA5000 CO₂ backup alarm system and the possibilities for your laboratory or organization.

Click here to download the brochure (PDF)

Additional information:

-Dimensions and weight:
Controller: 12.7 cm wide × 10.8 cm high × 4.45 cm deep, 0.45 kg
BVM Box: 15.88 cm wide × 12.7 cm high × 25.4 cm deep, 6.35 kg
Controller Alarm Box: 12.7 cm wide × 10.8 cm high × 4.45 cm deep, 0.91 kg

-Power supply and battery:
Input adapter: 100-240 VAC (fluctuation max ±10%), 50-60 Hz, max 1A
Output adapter: 15 VDC, 2.4A, 36 W
Battery: 12 VDC, 12 Ah, sealed lead acid, rechargeable, typically >48 hours backup

-Temperature management:
Temperature range: -200 °C to +50 °C
Standby temperature range: -70 °C to 0 °C

-Sensors and relays:
Sensor: Type T thermocouple, 3 meter cable
Output relays: SPDT contact, normally open (NO) or closed (NC), max 30 VDC / 1A

-CO2 management and safety:
Backup duration: >10 hours with a 23 kg CO2 cylinder
CO2 pressure: <1000 psi at 20 °C
Freezing switch: door must be closed for CO2 injection

-Material housing:
Rugged and durable bakelite for alarm box

-Mounting:
Flexible installation with compact dimensions

-Energy consumption:
Efficient with minimal power consumption thanks to advanced electronics

-Certification:
CE compliance, UL-listed power supply

-Warranty:
1 year full warranty

-Service and maintenance:
Recommended once a year
Backup battery should be replaced every 5 to 6 years mandatory